PVCHECK
EN - 38
6.3.4.1. Anomalous situations when performing IVCK tests
1. In case the instrument detects a voltage higher than
1000V at terminals P-N, P-E and N-E, it does not carry
out the test, gives out a long sound and displays the
message “Vin > 1000”.
15/05/12 15:34:26
Module : SUNPWR318
Vdc : 0.0 V
Irr : 0 W/m2
Tc : Auto °C
Voc,Isc:
Ri(1000V) - - -
M
Rpe (Cal) - - -
Vin > 1000
Selection
IVCK
2. In case the instrument detects a voltage lower than 15V
at terminals P and N, it does not carry out the test, gives
out a long sound and displays the message “Low
voltage”.
15/05/12 15:34:26
Module : SUNPWR318
Vdc : 0.0 V
Irr : 0 W/m2
Tc : Auto °C
Voc,Isc:
Ri(1000V) - - -
M
Rpe (Cal) - - -
Low voltage
Selection
IVCK
3. In case the instrument detects a voltage higher than > 5V
at terminals E and C, it does not carry out the test, gives
out a long sound and displays the message “Voltage >
Lim”.
15/05/12 15:34:26
Module : SUNPWR318
Vdc : 0.0 V
Irr : 0 W/m2
Tc : Auto °C
Voc,Isc:
Ri(1000V) - - -
M
Rpe (Cal) - - -
Voltage > Limit
Selection
IVCK
4. In case the instrument detects an Isc current higher than
10A, it does not carry out the test, gives out a long sound
and displays the message “Isc current too high”.
15/05/12 15:34:26
Module : SUNPWR318
Vdc : 0.0 V
Irr : 0 W/m2
Tc : Auto °C
Voc,Isc:
Ri(1000V) - - -
M
Rpe (Cal) - - -
Isc current too high
Selection
IVCK
Kommentare zu diesen Handbüchern