Ht-instruments SIRIUS87 Bedienungsanleitung Seite 15

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 88
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 14
SIRIUS 87
EN - 13
4.1.2.1. Mode "AUTO"
At the end of the test, if the average
resistance value Ravg results to be
lower than 5 the instrument emits a
double sound signal indicating the
positive outcome of the test and
displays one screen similar to the
screen alongside.
CAL LOW 
1.07
219 mA
AUTO
SAVING: The test can be stored pressing the SAVE key twice
(according to § 5.1).
4.1.2.2. Mode "R+TIMER" or “R-TIMER"
If during the test a resistance value
R+Timer or R–Timer lower than 5
was detected, the instrument (after the
set time has elapsed) emits a double
sound signal indicating the positive
outcome of the test displays one screen
similar to the screen alongside.
CAL LOW 
1.08
219 mA 5 s
R+
R-
T
IMER
SAVING: The test can be stored pressing the SAVE key twice
(according to § 5.1).
Average test
current value
Iavg.
Average
resistance value
Ravg.
Value of the test
current I+ or I-.
The symbols R+ or R- are displayed.
Duration time of
the test.
SAVE
SAVE
Seitenansicht 14
1 2 ... 10 11 12 13 14 15 16 17 18 19 20 ... 87 88

Kommentare zu diesen Handbüchern

Keine Kommentare