5.4.2. DIELECTRIC DISPLAY EXPLANATION
Adjustable/selectable parameters:
MODE - operation mode MANUAL, RAMP or BURN
UTEST NOM - nominal test voltage 250 5100 V AC
LIMIT - current limit 1 110 mA
CHAR - current character IAPP or IREAL
Figure 13: Display with DIELECTRIC test result
1 ..... Selected function.
2 ..... Progress bar, it follows test time during the measurement (in RAMP mode only).
3 ..... Test voltage applied during the measurement.
4 ..... Measurement screen touch-screen key.
5 ..... MODE touch-screen key to select operation mode (MANUAL, RAMP or BURN). Currently selected
mode is displayed on the bottom of the key.
6 ..... UTEST NOM touch-screen key to select nominal test voltage (250 up to 5100 V). Currently
selected value is displayed on the bottom of the key.
7 ..... LIMIT touch-screen key. Currently selected limit leakage current (trip out current) is displayed on
the bottom of the key.
8 ..... CHAR (character) touch-screen key to select the character of displayed leakage current (IAPP or
I
REAL). Currently selected character is displayed on the bottom of the key.
9 ..... Leakage current in green colour if the result is lower than or equal to set limit value. If break-
through occurred during the test then limit value will be displayed in red colour.
10 ... Real time clock (hh.mm.ss).
11 ... Unit of the test result (mA).
12 ... Set measurement time (in RAMP mode only).
13 ... Measurement result status (symbol
in green colour - result lower than or equal to set limit
value, symbol in red colour - break through occurred during the test or result higher than set
limit value).