6. RESET OF THE INSTRUMENT AND DEFAULT PARAMETERS
BEFORE OF A RESET PROCEDURE DOWNLOAD THE STORED TESTS ON A PC
6.1. RESET PROCEDURE
1. Press at the same time the keys DISP, CLR, RCL and then ignite the instrument.
2.
The screen nearby is displayed
for 5 seconds, after which the
instrument emits a sound signal
and then displays the screen
relative to the function selected
with the rotary switch.
es
ATTENTION: the reset procedure will cancel all the stored tests and it will set the
default parameters in the instrument.
6.2. DEFAULT PARAMETERS
After a RESET the default parameters set on the instrument are:
Function Parameter RESET default parameter
LOW
Mode AUTO
Calibration Offset
0
Mode R+/R- TIMER
TIMER is set at 1s
M
Mode MAN
Test voltage 500V
Mode TIMER TIMER is set at 60s
RCD type
A/AC
Mode x1
Test differential current 30mA
Contact voltage limit 50V
LOOP Z
S
/I
K
Mode P-N
EARTH ρ
Parameter DIST DIST = 1
Memory
Parameter PLA P = 1
Memory state 0